Test properties of Sp02 for detection of hyperoxia in sick infants*

Upper limit for SpO2SensitivitySpecificity+LR* −LR*
*Positive/negative LRs (likelihood ratios) calculated from data in article.
94%1.000.271.370
95%1.000.401.670
96%0.800.591.950.34
97%0.420.925.250.63
98%0.110.860.791.03
Individualised0.750.642.080.39